Konduktiv.mätare portabel CM35+ m väska exkl givare

Artikelnr: 43403502 Kategori:

Logga inRegistrera

  • Namn

  • Kontakt Info

  • Om dig

    • Leveransadress

    • Faktureringsadress

Beskrivning

Beskrivning

Portable condictivity meter CM 35

Ergonomic and robust instrument with IP67. Measures conductivity, salinity, total dissolved solids (TDS) and temperature. The range of cells covering a wide range of applications from clean water to viscous samples.

Simultaneous reading
It´s backlit LCD display allows simultaneous reading of the specific conductivity (salinity or TDS) to the selected reference temperature, the actual temperature of the sample and the duration of the measure.

Automatic change of scale
For reading with the highest resolution

Common advantages of portable CRISON
 • Maximum simplicity
• Light and ergonomic instruments, small size.
• Very resistant, designed to work in adverse conditions.
• Two measuring modes: by stability or continuous
• LCD display with icons and backlit display .
• Convenient carrying case for working in the field and in the laboratory.
• Functional. Software designed to greatly simplify the operation.
• Very easy to calibrate.
• Low Power Electronics.
• Intuitive keyboard for professional use.
• Connector for sensors of the highest quality. IP 67 tightness.
• Very competitive pricing when compared with similar instruments.
• 5 year warranty on all instruments.

Measurement Scales  (the resolution depends on the scale)
conductivity: 0.01 S / cm … 500 mS / cm
Salinity: 0.0 … 1999 mg / l NaCl 
                2.0 … 50.0 g / l NaCl
TDS: 0 mg / l … 500 g / l
Temperature : -20.0 … 150.0 ° C (-4.0 … 302.0 ° F)

Measurement error (± 1 digit)
Conductivity: ≤ 0.5%
Salinity: ≤ 0.5%
TDS: ≤ 0.5%
Temperature: ≤ 0.2 ° C (0.4 ° F)

Reproducibility  (± 1 digit)
Conductivity: ± 0.1%
Salinity: ± 0.1% l
TDS: ± 0.1%
Temperature: ± 0.1 ° C (0.1 ° F)

Temperature Compensation
Manual or through integrated Pt1000 temperature sensor

Reference temperature (RT)
20 or 25 ° C (68 and 77 ° F). Factory setting: 25 ° C (77 ° F)

Temperature coefficient (TC)
0.00 to 9.99% / ° C. (Default: 2% / ° C.)

TDS conversion factor
0.00 to 4.44. (Default: 0.64)

Cell constant accepted
between 0.05 and 50 cm-1. (Default: 1.0 cm-1)

Calibration of conductivity (EC)
with 1, 2 or 3 patterns to choose from. 147 mS / cm, 1413 uS / cm, 12.88 mS / cm and 111.8 mS / cm (25 ° C)
Manual calibration at any value.
theoretical calibration (factory setting)

Measurement modes
for stability and continuous.

Connector
MP-5, 5-pin multipin.

Display
Liquid crystal backlit with pictograms.

Keyboard
Membrane, 6 keys.

Environmental conditions
Operating temperature: 0°C to 50°C for
Storage Temperature: -15°C to 65°C for
80% relative humidity (non-condensing).

Low Voltage and EMC Directives
According to EC, EN 61010-1 and EN-61326-1.

Power Management
Auto Off after 5 minutes of inactivity.

Power
3 x 1.5V AA. Autonomy of over 400 hours.

Container
Material ABS housing. Degree of protection IP 67.

Resetting the temperature
correction probe deviation 25°C and 85°C CAT

Dimensions 186 x 73 x 38 mm
Weight 300 g

Including
– With 147 µS/cm, 1413 µS/cm, 12.88 mS/cm standard solution
– Calibration flasks
– Carrying case
Without cell

Suitable  cells: 5060,5061, 5062, 5063, 5064