Description
Benchtop autoclave
The ‘Autofill’ version has an integral reservoir and an in-built water re-circulation system. Once the reservoir is filled, this provides water for up to 15 cycles, and also acts as a collection vessel for condensate and the water that is automatically returned from the chamber at the end of each cycle, minimising water usage.
Format: Front loading
Capacity: 33 l
Temperature range: 100-138°C
Pressure range: 0.2-2.4 bar
Power requirements (heaters in chamber):
Single phase version: 230V, 13A, 50/60Hz, (N+E)
Three phase version: Not applicable
Power requirements (with optional steam generator):
Three phase only: Not applicable
Water requirements: Tap/softened water with <50ppm TDS; pH neutral. Autofill.
Drainage requirements: Not applicable
Air requirements: Not applicable
Vent/safety valve: No drain connection required
Chamber diameter: 346 mm
Chamber depth: 355 mm
Chamber usable depth: 294 mm
Required bench depth: 670 mm
Cooling locks: In accordance with H.S.E. PM73 preventing opening of the autoclave above 80ºC. (for fluid & discard cycles)
Alarms: For Cycle Fault – Cycle Interruption – Sterilize Failure – Water Low – Door Unlocked
För prisuppgift vänligen kontakta oss